The filter provides position-insensitive air removal and fluid delivery without operation.
TPN and lipid solution; drug therapy; apheresis solution
0.2 µm PES membrane is retentive of B. diminuta and meets USP 25/NF20 requirements for a sterilizing grade filter per modified ASTM F838-83 test methods
Maximizes delivery of critical drugs with proven low drug binding Supor membrane
Contains no exogenous surfactants, protecting patient from delivery of unwanted extractables
Colored housings and pad printing available for ease of final product identification
Materials of construction contain no natural rubber latex, or latex derivatives, PVC, or hydroxyvinyl chloride
Biological Safety | Materials of construction pass United States Pharmacopoeia (USP) Biological Reactivity Test, In Vivo |
---|---|
Bubble Point | Minimum (Imperial)46 psi |
Effective Filtration Area | 10 cm2 |
Filter Media | polyethersulfone membrane |
Flow Rate Range | (36 in. head w/0.105 in. tubing) Minimum 20 mL/min |
Inlet Connector | Standard 1.40 mm (0.055 in.) |
Length | 66 mm |
Material | (housing) Modified acrylic PTFE (Vent) |
Medical Device Registration | No |
Outlet Connector | Standard 3.43 mm (0.135 in.) |
Pore Dimensions | 0.2 µm |
Positively Charged Membrane | No |
Pressure Range | Maximum (Imperial)45 psi |
Pyrogenicity | < 0.25 EU/mL using the LAL test method |
Quantity | 2000 pkg |
Shipped Sterile | No |
Sterilization | Gamma and EtO sterilization |
Volume | (Priming)2.4 mL |
Width | 33 mm |